NG-HDT – HDT/Vicat Testing Machine for Plastics

Trust
  • Front view of the NextGen NG-HDT HDT/Vicat (VST) testing machine with an enclosed test chamber and touchscreen control panel
  • Inside view of the NextGen NG-HDT HDT/Vicat testing machine with a multi-station test rack, loading weights, and measurement probes arranged across the test stations
  • Close-up of the NextGen NG-HDT HDT/Vicat test station with specimen supports and probe tips for HDT deflection and Vicat softening temperature testing
  • HDT deflection-under-load setup on the NextGen NG-HDT tester, with a bar specimen resting on the support span and a loading nose positioned above the sample
  • Close-up of the immersion oil bath and heating elements inside the NextGen NG-HDT HDT/Vicat (VST) testing machine
  • Dimension drawing of the NextGen NG-HDT HDT/Vicat testing machine with front and side views
  • Accessory tool kit for the NextGen NG-HDT HDT/Vicat tester, including pliers, specimen cutters, tweezers, hex key, and calibration blocks

Standards

Description

The NG-HDT HDT/Vicat testing machine is designed to measure heat deflection temperature under load (HDT) and Vicat softening temperature (VST) for thermoplastics, including pipes and fittings. It runs these methods under a constant load with a controlled linear heating rate, which fits routine QA/QC work and material development.

The platform is offered in 3-station and 6-station configurations and uses a built-in touchscreen industrial PC to run the test sequence from loading and measurement through unloading and cooling. Measurement is based on Class A PT100 temperature sensing and LVDT displacement measurement, with an oil bath rated up to 300°C. The bath design includes turbulence suppression and cyclic heating functions to support steady heating conditions during the ramp.

Test data can be reviewed, searched, and exported over Ethernet. The system can also be paired with options such as a chiller for faster cooldown, an external oil-vapor extraction and filtration unit for high-temperature operation, and nitrogen protection to help reduce silicone oil consumption in high-usage labs.

```


Materials and Specimen Types This Machine Supports

This system is used for thermoplastic materials, including plastic pipes and fittings, where HDT and Vicat (VST) values are part of product qualification, incoming inspection, or material comparison work. It supports both test types on the same platform, so labs can cover common thermal performance checks without switching equipment.

Supported Specimen Sizes

For Vicat (VST), the maximum specimen size is 50 × 16 × 6.5 mm. For HDT, the maximum specimen size is 5.1 × 0.6 × 0.5″ (130 × 15 × 13 mm). A long-span option is available for ISO 75-3 / GB/T 1634.3 methods and supports specimens up to 10.2 × 0.6 × 0.5″ (260 × 15 × 13 mm), with a maximum span of 8.3″ (210 mm).

Vicat Specimen Size50 × 16 × 6.5 mm max.
HDT Specimen Size5.1 × 0.6 × 0.5″ (130 × 15 × 13 mm) max.
Long-Span Option10.2 × 0.6 × 0.5″ (260 × 15 × 13 mm) max.

Standards and Test Methods You Can Run (ISO / ASTM / GB/T)

ASTM, ISO, and DIN standards logosNG-HDT supports a broad set of HDT and Vicat (VST) methods used in plastics testing, including the most commonly referenced standards in North America and internationally, such as the ones listed below:

  • ASTM D648: Standard Test Method for Deflection Temperature of Plastics Under Flexural Load in the Edgewise Position - North American reference method for HDT. Used to compare how plastics deform under a specified flexural stress while temperature increases.
  • ASTM D1525: Standard Test Method for Vicat Softening Temperature of Plastics - North American reference method for VST. Measures the temperature at which a specified needle penetration occurs under defined load and heating conditions.
  • ISO 75-1: Plastics - Determination of temperature of deflection under load - Part 1: General test method - The general ISO framework for HDT style testing, including the overall procedure and common test setup.
  • ISO 75-2: Plastics - Determination of temperature of deflection under load - Part 2: Plastics and ebonite - The commonly used ISO part for most plastics, defining standard condition sets.
  • ISO 75-3: Plastics - Determination of temperature of deflection under load - Part 3: High-strength thermosetting laminates and long-fibre-reinforced plastics - Used when the material category and setup fall under high-strength laminates or long-fibre reinforced plastics.
  • ISO 306: Plastics - Thermoplastic materials - Determination of Vicat softening temperature (VST) - The ISO Vicat method, including standard load and heating-rate condition sets.
  • ISO 2507 (series): Thermoplastics pipes and fittings - Vicat softening temperature - Vicat test method and conditions written specifically for pipe and fitting samples.
  • GB/T 1634.1, GB/T 1634.2, GB/T 1634.3 - GB/T HDT method references aligned to temperature of deflection under load workflows.
  • GB/T 1633 - GB/T Vicat softening temperature reference for thermoplastic materials.
  • GB/T 8802 - Vicat softening temperature reference for thermoplastic pipes and fittings.

Automated Test Sequence: Loading, Heating, Measurement, Unloading, Cooling

Inside view of the NextGen NG-HDT HDT/Vicat testing machine with a multi-station test rack, loading weights, and measurement probes arranged across the test stationsNG-HDT is built around a hands-off test sequence that follows the same order labs run every day: bring the bath to the starting condition, apply the required load, ramp temperature at the selected rate, record displacement or penetration, then unload and cool down. For HDT, the setup uses the support span and loading nose to track deflection under flexural stress. For Vicat, the needle applies the specified load and the system tracks penetration as temperature increases. Once a run is complete, the unit moves into the cooling step so the next test can be set up.

Preset Standards + Custom Methods (Touchscreen Control)

The built-in touchscreen industrial PC lets operators select preset methods based on common ISO, ASTM, and GB/T standards, or create custom methods when a lab needs a specific combination of load and heating rate. During operation, the system monitors key checks at startup such as oil level, sensor status, and door-lock state, and it assumes control of the run without manual step-by-step intervention. Test data can be saved for later review, searched as history, and exported over Ethernet when the unit is connected to a network.

Method Family Built-In Preset Examples Typical Use
Vicat / VST ASTM D1525, ISO 306, GB/T 1633, GB/T 8802, ISO 2507 Softening-temperature testing by needle penetration under defined load and heating conditions.
HDT / DTUL ASTM D648, ISO 75-2, ISO 75-3, GB/T 1634.2, GB/T 1634.3 Deflection-temperature testing under flexural load using the required support span, loading nose, and stress level.
Custom Methods User-defined load, heating rate, and endpoint settings Useful when internal lab procedures, customer specifications, or special material programs require non-default settings.

Measurement Accuracy and Temperature Control

NG-HDT uses a Class A PT100 temperature sensor package and an LVDT displacement sensor system to capture the two signals that define HDT and Vicat testing: bath temperature and specimen movement. Temperature control is tied to the oil-bath layout, turbulence suppression, cyclic heating, and control parameters so the bath remains steady during the ramp, while the displacement system tracks deflection or penetration without relying on operator interpretation.

Displacement Resolution

0.001 mm resolution supports precise endpoint detection for HDT deflection and Vicat penetration measurements.

LVDT Measurement

The displacement system provides better than 0.01 mm accuracy across a 15 mm measurement range.

Class A PT100 Sensing

Class A PT100 temperature sensing supports controlled thermal ramping up to the 300°C bath range.


Cooling Options, Oil-Vapor Control, and Oil-Consumption Reduction

NG-HDT uses an immersion oil bath and supports water cooling as the standard cooling method. For labs that need shorter turnaround between runs, the system can be configured with an external chiller to speed up cooldown after a test. The platform also supports optional nitrogen protection on the oil surface, which helps reduce silicone oil consumption over time in higher-temperature, high-usage workflows.

With a chiller installed, typical cooling time is 45 minutes from 200°C down to 23°C. For laboratories running frequent high-temperature tests, the optional oil-vapor purifier can be used to extract and filter oil vapor from the test area.

Cooling MethodWater cooling, chiller option
Cooling Time45 min from 200°C to 23°C with chiller
Lab Environment OptionsOil-vapor purifier, nitrogen protection

Safety Systems That Prevent Overheat, Spills, and Damage

NG-HDT is designed around a closed test chamber that separates operators from the hot oil bath during a run. The protective shield stays closed for testing, and the door is locked while the program is active to reduce the chance of accidental contact or splashing. The system also checks basic conditions at startup, including oil level, sensor status, and door-lock state, and flags abnormal conditions before the test proceeds.

Closed Shield + Door Lock

A fully enclosed protection shield is used for testing, and the chamber door locks automatically during operation.

Overheat + Oil-Level Protection

Dual overtemperature protection, oil-level monitoring, and overflow protection help reduce the risk of heater damage or oil spill issues.

Lift + Outlet Safeguards

Lift obstruction protection, a relief valve, and a lockable oil outlet valve support safer operation during routine lab use.


Designed for Fast Cleaning and Easy Calibration

Close-up of the immersion oil bath and heating elements inside the NextGen NG-HDT HDT/Vicat (VST) testing machineThe test area is set up for routine lab maintenance, with a modular specimen rack that can be removed without tools. This makes it easier to clear fragments if a specimen drops and to keep the chamber clean between runs. The open-type oil tank uses quick-release knobs and can be cleaned in about five minutes. A dedicated calibration interface is built in to support quick checks with an external thermometer, without improvised setups.

To protect contact parts during day-to-day handling, the Vicat needle is kept suspended when not in use, and separating pliers are included to help remove specimens without damaging the needle or pin.

Maintenance note: The modular rack, quick-release oil tank, calibration port, and separating pliers are especially useful for labs that run repeated thermal tests and need fast chamber cleanup between batches.


Models and Technical Specifications

In the table below, you’ll find the available model configurations and the full technical specifications for the NG-HDT HDT/Vicat testing machine, including key performance parameters, supported test setups, utilities, and dimensions.

Station Configurations3-station and 6-station
Temperature RangeAmbient to 300°C
Control Interface10.4 in industrial touchscreen PC
Specification NG-HDT300 NG-HDT600
Test stations 3 6
Temperature range Ambient to 300°C (572°F)
Heating rate 120 ± 10°C/h (216 ± 18°F/h)

50 ± 5°C/h (90 ± 9°F/h)

Displacement resolution 0.001 mm
Displacement accuracy Better than 0.01 mm
Displacement measurement range 15 mm
HDT test span 2.52 in (64 mm) (GB/T 1634.2, ISO 75-2)

3.94 in (100 mm), 4.00 in (101.6 mm) (ASTM D648, optional)

HDT stress 1.80 MPa, 0.45 MPa, 8.00 MPa (GB/T 1634.2)

0.455 MPa, 1.82 MPa (ASTM D648) Custom (GB/T 1634.3, ISO 75-3)

VST load 10 ± 0.2 N, 50 ± 1.0 N
VST maximum specimen size 1.97 × 0.63 × 0.26 in (50 × 16 × 6.5 mm)
HDT maximum specimen size 5.12 × 0.59 × 0.51 in (130 × 15 × 13 mm)

9.06 × 0.50 × 0.28 in (230 × 12.8 × 7 mm) (optional for GB/T 1634.3, ISO 75-3)

Bath volume 4.76 gal (18 L)
Temperature sensor PT100
Cooling time 45 min (200°C to 23°C / 392°F to 73°F) with chiller
Cooling method Water cooling
Loading type Electric
Communication port LAN + 2 × USB 2.0
Power consumption 3.5 kW
Power supply 1-phase, 220 V AC ± 10%, 50 Hz ± 2%
Weight 353 lb (160 kg)
Dimensions 50.4 × 20.9 × 31.5 in (1280 × 530 × 800 mm) (L × W × H) 50.4 × 24.4 × 49.6 in (1280 × 620 × 1260 mm) (L × W door-open × H door-open)

Dimension drawing of the NextGen NG-HDT HDT/Vicat testing machine with front and side views


Standard Configuration and Optional Accessories

Below is an overview of what ships with the NG-HDT system as standard, plus the optional add-ons that can be supplied based on your lab setup. Quantities for station-specific items differ between the 3-station and 6-station configurations.

Standard Configuration

This table lists the components supplied with the system. Items marked 3/6, 4/7, etc. depend on whether the unit is configured with 3 or 6 test stations.

Name Description QTY
Test stations Configured by model 3/6
Support span 2.52 in (64 mm) (ISO 75)

4.00 in (101.6 mm) (ASTM D648, Method A) 3.94 in (100 mm) (ASTM D648, Method B)

3/6
Immersion bath Oil bath assembly 1
Temperature sensor PT100, Class A 4/7
Displacement sensor LVDT displacement measurement 3/6
Computer Built-in 10.4 in industrial PC, capacitive touch screen 1
HDT loading nose R = 0.12 in (3.0 ± 0.2 mm) 3/6
VST needle Cross-section area 1.000 mm² ± 0.015 mm² 3/6
Universal weight set Combined load up to 12.1 lb (5500 g).

All loads from 1 g to 5500 g are possible in 1 g steps.

3/6
Test software GenTest, English version 1
Test accessories Tweezers, specimen scissors, specimen separating pliers, specimen box, wrenches (1.5 mm and 2.5 mm), oil filling funnel, inlet and outlet water pipes (rated to 300°C / 572°F), 78.7 in (2 m) length 1
Calibration accessories Gauge blocks (2 mm, 3 mm, 2 × 4 mm)

Temperature calibration hole (Φ 3.2 mm)

1
Quality certificate Included 1
Operation manual USB drive 1

Accessory tool kit for the NextGen NG-HDT HDT/Vicat tester, including pliers, specimen cutters, tweezers, hex key, and calibration blocks

Optional Accessories

Optional items can be supplied to match your workflow, cooling requirements, or test method needs.

Name Description
Universal weight set Combined load up to 12.1 lb (5500 g). All loads from 1 g to 5500 g are possible in 1 g steps.
Vicat weight set 10 N and 50 N in accordance with ISO 306 and ASTM D1525 (one set required per measuring station).
HDT loading nose R = 0.12 in (3.0 ± 0.2 mm).
VST needle Cross-section area 1.000 mm² ± 0.015 mm².
ASTM support roller Optional support roller setup for ASTM D648 configurations when required by the selected method setup.
Work bench 55.1 × 25.6 × 29.1 in (1400 × 650 × 740 mm).
Chiller 2.2 kW external cooling option for faster cooldown between test runs.
Oil-vapor purifier 80 W, 3-layer filter, 51.2 in (1.3 m) pipe, 16.7 × 9.8 × 16.1 in (425 × 250 × 410 mm).
Filter set (for oil-vapor purifier) First layer: 5 pcs. Second layer: 2 pcs. Third layer: 1 pc.
Nitrogen protection system Optional oil-surface protection for labs that want to reduce oil oxidation and silicone oil consumption during high-temperature operation.
Adjustable-span support Meets GB/T 1634.3 and ISO 75-3. Maximum span: 8.27 in (210 mm). Maximum specimen: 10.24 × 0.59 × 0.51 in (260 × 15 × 13 mm).

Have Questions or Need to Confirm a Detail?

If you’re not sure which method you need to run, what station count fits your workload, or whether you need options like a chiller, oil-vapor purifier, nitrogen protection, or an adjustable-span support, send us a quick request online. Share your material type, applicable standard, specimen size, and expected test volume, and we’ll get back to you with the right configuration for your lab.

For Laboratories

Need Additional Quality Control Equipment to Build Your Laboratory?

HDT and Vicat testing is only one part of a complete quality control laboratory. NextGen can help you equip your entire lab with universal testing machines, impact and falling weight testers, melt flow indexers, HDT and Vicat systems, hardness and durometer systems, environmental chambers, and other supporting solutions, all coordinated as one complete project.

  • One partner to equip your complete laboratory
  • Installation, training, calibration
  • Support for the life of the equipment

FAQs

On the bench, the NG-HDT measures 50.4 × 20.9 × 31.5 in (1280 × 530 × 800 mm), length by width by height. Allow room for the door as well: with it open, the envelope grows to 50.4 × 24.4 × 49.6 in (1280 × 620 × 1260 mm), so keep clear space above and in front of the unit. The machine weighs 353 lb (160 kg).

It runs on a single-phase 220 V AC supply (±10%, 50 Hz ±2%) with 3.5 kW power consumption. The installation also needs water inlet and outlet connections for the standard water cooling, and the pipes, rated to 300°C (572°F) and 78.7 in (2 m) long, ship with the machine. A LAN port and two USB 2.0 ports handle data connections.

If the lab has no suitable bench, we can supply a work bench of 55.1 × 25.6 × 29.1 in (1400 × 650 × 740 mm). Installation and training come with our lab projects, so talk to our team about the room before delivery and we will check space, power and water with you.

HDT and Vicat results show how a plastic behaves as it heats up under load, but a qualification file rarely stops there. When we equip a plastics lab, the thermal system is planned alongside a melt flow indexer for flow behaviour, universal testing machines for tensile and flexural properties, and impact and falling weight testers for toughness.

Hardness and durometer systems and environmental chambers are part of the same lab projects, so a single plastics lab can cover thermal, mechanical, flow and hardness checks under one roof.

Because all of these come from NextGen, the lab works with one partner for installation, training, calibration and support for the life of the equipment. See how we equip complete laboratories, or send us the standards your material program has to meet and we will map the equipment to them.

For Vicat softening temperature, the NG-HDT works to ASTM D1525, ISO 306, the ISO 2507 series for thermoplastic pipes and fittings, GB/T 1633 and GB/T 8802. All five are stored as preset methods on the touchscreen.

For heat deflection temperature, it covers ASTM D648 and ISO 75 Parts 1, 2 and 3, together with GB/T 1634.1, GB/T 1634.2 and GB/T 1634.3. ASTM D648, ISO 75-2, ISO 75-3, GB/T 1634.2 and GB/T 1634.3 are preset. ISO 75-3 and GB/T 1634.3 apply to high-strength thermosetting laminates and long-fibre-reinforced plastics, and they run with a custom stress value and the optional adjustable-span support.

Each method still depends on the edition, the specimen and the condition set your test plan names. If a purchase specification cites one of these documents, include the edition and the condition set with your enquiry and we will match the spans, loads and weights to it.

Each test station has its own LVDT displacement sensor, so the NG-HDT300 carries three and the NG-HDT600 carries six. In an HDT run the sensor tracks deflection of the bar under the loading nose, and in a Vicat run it tracks penetration of the needle, while the bath temperature rises.

Resolution is 0.001 mm and accuracy is better than 0.01 mm over a 15 mm measuring range. That margin matters at the end of a test, because the result is the temperature at which a specified deflection or penetration is reached, and a fine reading lets the system catch that point as it happens.

The endpoint is detected by the instrument, not read off a dial by an operator, which removes one source of difference between operators and shifts. Displacement and temperature are stored with each test, so earlier runs can be searched and reviewed later.

Two linear heating rates are built in: 120 ± 10°C/h (216 ± 18°F/h) and 50 ± 5°C/h (90 ± 9°F/h). The selected rate is held while the specimens stay under constant load, and the temperature at which each one reaches its endpoint is the test result.

The two rates are the ones the common Vicat and HDT condition sets are built on. ISO 306, for example, groups its conditions as A50, B50, A120 and B120, each a combination of load and heating rate, so both rates are needed on the same machine.

The rate is set by the method chosen on the touchscreen, so nobody adjusts heating by hand during the ramp. Once the endpoint is reached, the run moves on to unloading and cooling without operator steps, which keeps the procedure identical from batch to batch.

The NG-HDT measures two temperatures that describe how a thermoplastic softens as it heats under load: heat deflection temperature (HDT) and Vicat softening temperature (VST). Both tests run on the same platform, so a lab covers them without a second instrument.

The materials are thermoplastics, including plastic pipes and fittings. Labs use the results in product qualification, incoming inspection and material comparison, which puts the machine in routine QA/QC work as well as in material development.

A run is automatic from start to finish. The system loads the specimens, ramps the bath at the selected rate, records deflection or penetration, then unloads and cools so the next batch can go in. With three or six stations, several specimens are tested in one heating cycle, and results can be reviewed, searched and exported over Ethernet.

The oil bath runs from ambient temperature up to 300°C (572°F) and holds 4.76 gal (18 L). Specimens are tested in this immersion bath, so the oil carries heat to every station during the ramp.

Holding the ramp steady across that span depends on the bath design. The NG-HDT bath uses turbulence suppression and cyclic heating, which keep heating conditions steady while the temperature climbs through the test.

Cleaning is quick. The open-type oil tank is released with knobs and can be cleaned in about five minutes, and the modular specimen rack lifts out without tools, so fragments from a dropped specimen can be cleared between runs.

There are two models. The NG-HDT300 has three test stations and the NG-HDT600 has six. Temperature range, heating rates, displacement measurement, bath, power, weight and footprint are the same on both, so the choice comes down to how many specimens you test in each heating cycle.

Station-specific parts scale with the count. Each station has its own support span, HDT loading nose, Vicat needle, LVDT sensor and universal weight set, and the machine carries four temperature sensors on the three-station model and seven on the six-station model. The optional Vicat weight set is ordered one per station.

Pick the count against your expected volume: the six-station model doubles the specimens per run, which matters when HDT and Vicat batches queue up. Request a quote for either model with your weekly test volume and standards, and we will price the matching station kit.

Both tests heat a specimen at a controlled rate under a constant load. They load it differently, though, and so they report different temperatures.

In HDT, a bar sits edgewise on a support span and a loading nose applies a set flexural stress while the bath heats. The result is the temperature at which the bar reaches the specified deflection. On the NG-HDT the stress is set to 1.80, 0.45 or 8.00 MPa for ISO 75-2 and GB/T 1634.2, or to 0.455 or 1.82 MPa for ASTM D648.

In Vicat, a needle presses on the specimen under a 10 N or 50 N load, and the result is the temperature at which it reaches the specified penetration. Vicat is the method written into the pipe and fitting standards, such as GB/T 8802 and the ISO 2507 series.

Our article HDT and Vicat: Two Softening Temperatures That Answer Different Questions sets the two methods side by side.

Every NG-HDT ships with its three or six test stations. Each station is fitted with a support span, an HDT loading nose of R 3.0 ± 0.2 mm, a Vicat needle of 1.000 ± 0.015 mm² cross-section, an LVDT displacement sensor and a universal weight set that combines up to 5500 g (12.1 lb) in 1 g steps.

The base unit includes the immersion oil bath, four or seven temperature sensors and a built-in 10.4 in industrial PC with capacitive touchscreen, running our GenTest software in English.

Handling tools are in the box: tweezers, specimen scissors, specimen separating pliers, a specimen box, 1.5 mm and 2.5 mm wrenches, an oil filling funnel and the inlet and outlet water pipes. For calibration you get gauge blocks of 2 mm, 3 mm and 2 × 4 mm, and the machine has a Φ3.2 mm temperature calibration hole for an external thermometer.

A quality certificate and the operation manual on a USB drive complete the delivery.

Yes. The touchscreen control system includes preset method options and supports custom methods for internal procedures or customer-specific requirements. This helps labs run routine standards-based testing while keeping flexibility for special material programs or non-standard evaluation work.

For mixed-method labs, request a quote with the required standards listed.

The system supports water cooling, and an optional chiller is available for faster cooldown. With the chiller option, the bath can cool from 392°F to 73°F (200°C to 23°C) in about 45 minutes, depending on operating conditions.

If faster turnaround is important for your lab, request a configuration quote with the cooling option included.

The system uses Class A PT100 temperature sensing. This supports stable temperature measurement during the controlled heating cycle, which is critical for repeatable HDT and Vicat results.

For labs comparing older equipment against current requirements, ask us to review the setup.

The standard configuration supports common HDT and Vicat specimen formats, including 2.52 in (64 mm), 3.94 in (100 mm), and 4.00 in (101.6 mm) support spans depending on the selected method. It also includes HDT loading noses, Vicat needles, gauges, and related tooling for routine operation.

If your specimens require a non-standard span or fixture setup, send us the details.

Optional accessories include universal weight sets, Vicat weight sets, automatic cooling, oil vapor extraction and filtration, nitrogen protection, computer options, and related items for specific lab requirements.

If you already know your standards, station count, and throughput target, request a detailed quote.

The NG-HDT includes a closed shield, door lock, dual overheat protection, oil-level monitoring, overflow protection, lift obstruction protection, pressure relief protection, and a lockable oil outlet. These features help protect the operator, specimens, oil bath, and mechanical system during heating, testing, unloading, and cooling.

For high-throughput labs, ask about safety and workflow configuration.

Related Products

NG-AGOV-ADV – High-Temperature Aging Oven

NG-AGOV-ADV – High-Temperature Aging Oven

NG-AGOV-ADV is a high-temperature aging oven for testing how rubber, plastics, elastomers, insulating materials, and similar products respond to long-term heat exposure. It simulates accelerated aging, thermal stability, and heat resistance conditions up to 300°C using forced convection, PID control, programmable cycles, stainless chamber construction, adjustable trays, over-temperature protection, and real-time monitoring. ISO 188, ASTM D573, IEC 60216-4-1, and GB/T 3512 support compliance workflows.

Read more
NG-EC Series – Environmental Test Chambers (100–1000 L)

NG-EC Series – Environmental Test Chambers (100–1000 L)

NextGen NG-EC Environmental Chambers provide controlled temperature and humidity conditions for material, product, and quality-control testing. Available in 100, 150, 225, 408, and 1000 L sizes, they feature insulated steel construction, stainless interior walls, mechanical compression refrigeration, circulating air ducts, adjustable sample racks, a viewing window, and 7-inch touchscreen control to support stable, repeatable environmental simulation.

Read more